top of page

Introduction

In order to achieve precise and trustworthy data, proper setup and calibration of the NIRScout is key. This process involves:

  • Conducting a phantom test prior to participant data collection

  • Selecting the desired optode montage for the cap and fitting the cap onto the participant

  • Calibration of the fNIRS system 

  • Collecting the data from the participant

 

​​Preparation of the NIRScout (prior to participant arrival)

 

At least 30 minutes prior to the arrival of the participant, switch on the NIRScout system (switch on the backside of the chassis) so that the red "Power" light is on. (Note: The green "Status" light may be on or flashing, but this does not mean the system is on.)

Perform a Phantom Test

A calibration test is used to ensure that all optodes are functioning properly. While a phantom test is not necessary to do before every participant's data is collected, it should be completed at least once before an entire study is conducted. 

 

Place optodes in chronological order, alternating source and detector (i.e. Source 1, Detector 1,

Source 2, Detector 2, etc.). There should not be a source next to another source or a detector next

to another detector. 

Place the Phantom setup inside of a dark sleeve (black laptop case) and close it as much as possible,

careful to not damage the cords. Turn the light in the lab room off and close the door as much as

possible. (NOTE: Don't completely shut the door; it will lock)

 

                      Open the NIRStar application on the desktop. (If you turn the system off and power it

                      back on again while NIRStar is running, you must shut down NIRStar and reopen it again

                      otherwise tests will not run properly.) Select "Configure Hardware" in the top left

                      corner of the screen. Select "Hardware Specification" at top. Under the "Study Type"

                      dropdown, select "Static Phantom". (NOTE: Available sources should be set to "8"

                      and Available detector cards should be "2") Select "OK".

 

Select the "CALIBRATE" button on the right hand side of the screen. The test is done when the loading screen disappears and "Calibration finished" is displayed in the System Status History window. If all boxes on the Signal Quality window are green (ignoring any empty grey spaces), the calibration was successful and you may move on to fitting the optodes onto a cap. If any squares are yellow, red, or white, select the "DETAILS" button above the calibrate button to decipher where the problem is.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Placing the NIRS Cap and Optodes​

There are various montages you can choose to use depending on where you are looking to measure on the participant's head. For the purposes of this explanation, the "Prefrontal_8x8" montage will be used. This montage setup is printed on the wall in the lab room. 

Once the participant arrives, welcome the participant and ask if they would have a seat in the EEG collection room. Begin by explaining the procedure, what you will be doing, and what the participant will be doing during the testing as well as the time requirement.

Measure their head with a tape measure from their nasion (area in between the eyebrows) to their inion (the projection of the occipital bone on the back of the skull) to determine which cap size is appropriate. Place the optodes in the correct montage setup on the cap with the wires facing the back of the cap. Be very careful with the optodes and wires, as they will easily scratch or crack, and the wires may easily be damaged by pinching. 

Carefully place the cap onto the participant's head ensuring that their ears align with the ear holes and the chin strap is snug but not tight. Move as much hair out of the way as possible. After the cap is on, an over-cap may be necessary if the participant cannot sit in the dark to perform the study. 

Watch this video for a demonstration on how to apply the NIRS cap:

PhantomSetup_edited.jpg

Phantom test setup. Begin with #1 source and detector at bottom left, adding optodes moving right.

nirstar application.png
image (1).png

Successful phantom test

 

Configuring NIRStar Settings for Data Collection

 

Open the NIRStar application on the desktop.

Select "Configure Hardware" in the top left corner of the screen.

Under "Predefined Montages" there is a "Montage" dropdown. Select the montage you will be using. The matching optode layout should be displayed. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Under "Hardware Specification", there is a "Study Type" dropdown. Select "Human Subject". (NOTE: Available sources should be set to "8" and Available detector cards should be "2") Select "OK".

 

 

 

 

 

 

 

 

 

 

 

Calibrating the NIRScout

Once the cap and over-cap are secure, turn off the lights, close the door and click “CALIBRATE” under “Main Functions”. After the system is calibrated, click “DETAILS” which will bring up the quality scale window, where you are able to view the topographical layout and operational values of the optodes. The numbers within the boxes describe the sensors and detectors. For instance, if a box read “3-2”, this would indicate sensor 3 and detector 2 (sensor is listed first, detector second) Within this screen, toggle between the tabs at the top of the screen to view how well the optodes are placed. Boxes displayed in green are optimal, and values should be compared to the quality scale bar on the right end of the window.

If any boxes appear yellow, red, or white, first determine which sensor/detector it correlates with. Then, carefully remove the optode from the cap and, using a small pick or cotton swab, move as much hair out of the way of where the optode sits as possible. ​Recalibrate to ensure that the problem areas turn green.

Collecting Data

 

Visualizing Data and Adjusting Skin Blood Flow Levels

 

Following the selection of the montage, adjust the skin blood flow waves by sliding “Scale Factor of each channel [mmol/l]” and “LP Filter Cutoff (Hz)” until the waves are more linear, as shown in the image below.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Before beginning the recording, click, “configure hardware” and within the Hardware configuration

window, click “Preview Display” This will open up a GUI that allows the visualization of blood flow

by hemisphere. Once the recording begins you will not be able to open this GUI. After the recording

has begun, click “Prepare display” within the GUI window to see real-time activity. 

 

You can also preview the data to ensure it appears accurate by selecting "PREVIEW". The data seen

here will not be saved until the "RECORD" button is selected. A "Subject Demographics" window will

be brought up. Fill in the information as desired and click “Done”.

To record the data, select "RECORD". Wait roughly 10-20 seconds before beginning any study protocol

so as to not miss any data.

Event triggers can be placed periodically into the data to mark moments of importance during

the study. This can be done manually by selecting one of the "Triggers" at the bottom of the

"Data Visualization" screen; however, there is much room for error via this method. If

conducting a formal study, utilize NIRStim or Presentation to automate the triggers. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

When completed, wait another 10-20 seconds before selecting "STOP". 

 

Thank the participant for their time and help them out of the cap.

image (2).png
Screenshot 2025-05-29 at 12.57_edited.jpg
bottom of page